CRT |
- Configuration
and useful screen :
6 inch rectangular
screen with internal graticule,8x10div.
(1div = 1cm) marking for
measurement of rise time, 2mm subdivisions
along the central axis.
- Acceleration potential : +1.9kv approx.(ref.
Cathode)
- Phosphor : P31
- Focusing : provided
- Trace rotation : provided
- Intensity control : provided |
 |
|
Z-AXIS INPUT
(INTENSITY
MODULATION) |
- Input
signal :
positive going signal decreases
+5Vp-p or more signal cases
intensity noticeable modulation
at normal intensity setting
- Bandwidth : DC to 2MHz(-3dB)
- Coupling : DC
- Input impedance : 25kз - 30kз typical
- Maximum input voltage : 30V(DC + peak
AC) |
 |
|
VERTICAL
DEFLECTION |
- DC coupled(-3dB)
:
DC to 20MHz normal / DC
to 10MHz magnified (CH 1 only)
- AC coupled(-3dB) :
10Hz to 20MHz normal /
10Hz to 10MHz magnified (CH 1 only)
- Modes :
CH1, CH2, ADD, DUAL(CHOP
: Time/div. switch from 0.2s to 1ms,
- ALT : Time/div. switch from 0.5ms
to 0.2レs)
- Deflection factor (normal): 5mV/div.
to 20V/div. in 12 calibrated steps
of 1-2-5sequence.
Continuously variable between
steps at least 1:2.5 X 5.
- Deflection factor (5MAG): 1mV/div.
to 4V/div. in 12 calibrated steps
(CH1 only).
- Accuracy : normal 【3%, magnified :
【5%(CH 1 only)
- Input impedance : approx. 1Mз in parallel
with 30pF
- Maximum input voltage : direct 400V
(DC + peak AC)
- Input coupling : AC, GND, DC
- Rise time : 17.5ns or less (35ns or
less in 5MAG)
- CH1 output : 25mV/div. into 50з /
DC to 10MHz(-3dB)
- Polarity inversion : CH2 only |
 |
|
HORIZONTAL
DEFLECTION
|
- Sweep
mode : NORM, X-Y, X 10, Variable
- Time base A : 0.2レs/div. to 0.2s/div.
in 19 calibrated steps of 1-2-5
sequence uncalibrated continuous
control between steps at least 1:2.5
- Sweep magnification : 10 times(maximum
sweep rate : 20ns/div.) 50ns/div.
or 20ns/div. 【10%
- Accuracy : 【3%(0∩ to 35∩), 【5%(0∩~10∩,
35∩~40∩) / additional error for
magnifier : 【2% |
 |
|
TRIGGER
SYSTEM |
- Modes
: AUTO, NORM, TV-V, TV-H
- Source : VERT(DUAL/ALT), CH1, LINE,
EXT
- Coupling : AC
- Slope : + or -
- Sensitivity and frequency in AUTO,
NORM 20HZ - 2MHz :
0.5div. or 2div.(VERT)
in INT / 0.2Vp-p in EXT
- Sensitivity and frequency in AUTO,
NORM 2Hz - 20MHz :
1.5div. or 3div.(VERT)
in INT / 0.6Vp-p in EXT
- TV-V, TV-H : at least 1div. or 1.0Vp-p
- External trigger input impedance :
1Mз 【 10%
- Maximum input voltage : 400V (DC +
AC peak) |
 |
|
X-Y
OPERATION |
- Sensitivity
: same as vertical deflection for both
X-axis(CH1) and Y-axis
(CH2) except accuracy 【
5% in X-axis
- X-axis bandwidth : DC to 500kHz (-3dB)
- X-Y phase difference : 3∑ or less
as DC to 50kHz |
 |
|
CALIBRATOR |
- Probe
adjustment :
approx. 1kHz frequency(20%),
0.5V(10%) square wave with 40-50% duty
ratio |
 |
|
POWER
&SUPPLY |
- Line voltage
range :
115V(98-125V) with 125V
1A fuse or 230V(198-250V) with 250V
0.5A fuse
- Line frequency : 50/60Hz
- Power consumption : approx. 45W |
 |
|
PHYSICAL
CHARACTERISTICS |
- Weight : 7.8kg
- Dimension : 316mm(W) X 143mm(H) X
406mm(L) |
 |
|
OTHERS |
- Accessories
including : Operation manual, Spare
fuse,Power cord
- Optional accessories (sold separately)